8 Hygienic reprocessing
1. Assemble the patient valve.
When doing so, note:
•
the side of the PEEP control diaphragm labeled "TOP" must
face upward toward the control cover.
•
the arrow on the control cover must point toward the
patient.
2. If available: Secure the protective cap strap on the patient
valve:
•
or
•
(only with reusable breathing circuits with flow
measurement
and
CO
2
measurement)
3. Connect the patient valve to the ventilation hose.
1
2