4 Preparation and operation
Depending on the preset in the operator menu, the device
switches to one of the following modes with the ventilation
parameters preset for the patient group
settings for emergency modes and ventilation modes",
page 257)
and shows a pressure gauge view:
•
IPPV
•
B ASB (only with activated B ASB option)
Result
An emergency mode for a particular patient group is activated.
4.7.2 Calling up the parameters of the patient last
ventilated
Requirement
The device is switched off.
1. Switch on the device.
After the self-test, the device displays the start menu:
If the device was switched off for < 30 seconds (e.g., for a
battery replacement), a countdown in the
Previous patient
field counts down 20 seconds.
If the device was switched off for > 30 seconds, the
countdown for the last patient is deactivated.