WM
6807
1
0
5
/2
021
MEDUMAT Standard
2
EN
39
3 Description
No.
Designation
Description
Patient valve
Switches between inspiration and expiration.
FlowCheck sensor connection line
with MEDUtrigger (only with Flow
measu ASB option)
Connects MEDUtrigger and the FlowCheck sensor to
the device. Alternatively, you can also connect the
FlowCheck sensor connection line or the connection
line of MEDUtrigger here.
Water filter
(only with Capnography option)
The water filter protects the measuring chamber of
the device against moisture and contamination from
the patient's respiratory gas.
Measuring circuit
The device measures the patient's vital parameters
via the measuring circuit. The measuring circuit
comprises:
•
Measuring circuit connector
•
PEEP control tube
•
Pressure measuring tube
•
CO
2
measuring tube
(only with Capnography option)
Measuring circuit connector
Connects the measuring circuit to the connection
for the measuring circuit on the device.
PEEP control tube
The device controls the patient valve and the PEEP
via the PEEP control tube.
CO
2
measuring tube
(only with Capnography option)
Measures the CO
2
content in the respiratory gas of
the patient.
Pressure measuring tube
Measures the ventilation pressure at the patient.
Y-piece
(only with Capnography option)
Connects the pressure measuring tube and the CO
2
measuring tube with the elbow of the breathing
circuit.
FlowCheck sensor connector (only with
Flow measu ASB option)
Connects one of the following connection lines to
the FlowCheck sensor:
•
FlowCheck sensor connection line
•
FlowCheck sensor connection line with
MEDUtrigger
Blanking plug
Closes the CO
2
connection
Elbow with CO
2
connection
•
Connects the rest of the breathing circuit to
the mask or endotracheal tube.
•
Enables the connection of the pressure measuring
tube and the CO
2
measuring tube (only with
Capnography option).
FlowCheck sensor (only with Flow
measu ASB option)
Measures the flow to the patient and to the device.
Summary of Contents for MEDUMAT Standard2
Page 1: ...MEDUMAT Standard2 Ventilator Instructions for Use for Devices from Software Version 5 1...
Page 269: ......
Page 270: ......
Page 271: ......