8 Hygienic reprocessing
Disruption or failure of therapy due to unsuitable cleaning
agents and disinfectants!
The use of incorrect cleaning agents and disinfectants can
cause the device to malfunction. This can result in serious or
life-threatening injury to the patient.
Never clean the device and accessories with bleach, bleach
solutions or phenolic compounds.
Only use the cleaning agents and disinfectants recommended
in these instructions for use
Loss of mechanical or electrical safety resulting from
reprocessing of the device and accessories with unsuitable
cleaning agents and disinfectants!
Using incorrect cleaning agents and disinfectants might cause
damage to the surface of the device and/or accessories, as well as
impairing electrical and insulating properties. This can result in
severe or life-threatening injury to the user and patient.
Only use the cleaning agents and disinfectants recommended
in these instructions for use
Disruption or failure of therapy due to liquid in the breathing
circuit after hygienic reprocessing!
Drips in the reusable measuring circuit or FlowCheck sensor can
distort the measurement results. This can result in serious or
life-threatening injury to the patient.
Following hygienic reprocessing of the breathing circuit,
allow all breathing circuit components to dry thoroughly.