74
EN MEDUMAT
Standard
2
WM
6807
1
0
5
/2
021
4 Preparation and operation
4. If necessary: Set the parameters of the ventilation mode.
5. Select the
start
field.
Result
The ventilation mode is changed.
4.7.6 Operating the device in oxygen or Air Mix mode
Requirement
•
The device is switched on.
•
A ventilation mode is set.
1. Briefly press the Air Mix button
.
Air Mix
appears in the status line and the device is operated in
Air Mix mode.
2. Briefly press the Air Mix button
:
•
If
100 % O
2
is set as the supply gas
100 % O
2
appears in the status
line and the device is operated with 100 % oxygen.
or
•
If
93 % O
2
is set as the supply gas
appears in the status line
and the device is operated with concentrator oxygen.
93%
O
2
Summary of Contents for MEDUMAT Standard2