10 Alarms and error messages
The functionality of an option is
not available
Option is deactivated in the
operator menu
Activate the option in the operator
menu
.
Option is not enabled in the
operator menu
Enable the option in the operator
menu with the enable code
Power failure/device failure:
•
Black screen
•
Alarm LED flashes
•
Audio alarm output
Battery empty and device not
connected to the line power
Check power supply.
Device defective
Switch off the device and have it
repaired.
The device switches off in NVG
mode
Battery empty and device not
connected to the line power
Check power supply.
Device defective
Switch off the device and have it
repaired.
Not possible to establish
Bluetooth
®
connection during
ventilation
Bluetooth
®
deactivated in user
menu
Activate Bluetooth
®
in user menu
Distance between two devices
too far
Reduce distance and move any
objects which might interfere.
Bluetooth
®
module of device or
external data communication
device is defective
Repair Bluetooth
®
module of
device or external data
communication device.
Devices are not paired
Pair devices in user or operator
menu
Unable to export files to SD card
SD card not formatted correctly
Use SD card WM 29791
Format SD card with FAT32 file
system
Fault
Cause
Remedy