2 Safety
Risk of infection if the device is used in a contaminated
atmosphere!
If the device is used in a contaminated atmosphere, it may suck
in contaminated or infected ambient air and harm the patient.
Only operate the device in a contaminated atmosphere with
a hygiene filter.
Risk of injury if the device is used in a dusty atmosphere!
If the device is used in a dusty atmosphere, it can suck in dust and
contaminants from the ambient air. Dust and contaminants may
reach the lungs of the patient and harm them.
Only operate the device with a hygiene filter.
Change the hygiene filter following operation in a very dusty
atmosphere.
Risk of explosion if the device is used in explosive
atmospheres!
Flammable gases and anesthetics may cause spontaneous
explosions and thereby bring about injury to the patient, user and
bystanders.
Do not use the device in combination with flammable gases or
anesthetic gases.
Risk of injury due to device or component malfunction!
A damaged device or damaged components may result in injury
to the patient, user or bystanders.
Only operate the device and components if they are externally
undamaged.
Only operate the device and components if the function check
has been successfully completed.
Only operate the device if the display is functional.
Keep an alternative ventilation unit at the ready.
Risk of injury if the pneumatic connections within the device
are closed off or blocked!
When oxygen is supplied via a central gas supply system which has
not been properly cleaned or is moist, the pneumatic connections
within the device may become blocked by contaminants or
particles or suck in moisture.
Only operate the device from the central gas supply system
which is clean and dry.