184
EN MEDUMAT
Standard
2
WM
6807
1
0
5
/2
021
8 Hygienic reprocessing
6. Rinse the pressure measuring tube and PEEP control tube with
drinking-water quality from the inside at least eight times with
a disposable syringe.
When doing so, note: Only rinse in one direction.
7. Allow the reusable measuring circuit to dry out completely.
8. Check the reusable measuring circuit for residue and any
remaining dirt.
9. If visible dirt remains: Repeat the manual cleaning.
Result
The reusable measuring circuit has been cleaned manually.
8.6 Wipe disinfecting parts
Requirement
The parts have been manually cleaned and are visibly clean
(see "8.5 Cleaning parts manually", page 181)
.
1. The parts approved for wipe disinfection can be found in the
cleaning and disinfection plan
2. The cleaning agents, doses and exposure time for the individual
parts can be found in the cleaning and disinfection plan
(see "8.11 Cleaning and disinfection plan", page 196)
3. Wipe-disinfect the parts with one of the agents listed
(see "8.11 Cleaning and disinfection plan", page 196)
When doing so, note:
•
Wipe down all surfaces with a cloth at least 8 times.