9 Function check
6. Check the alarm system:
•
The alarm light must flash red.
•
The device must emit at least one audible alarm in alarm
system test 1 and alarm system test 2.
7. If the alarm system is functioning: Press the navigation knob
yes
each time.
8. If the alarm system is not functioning: Press the navigation
knob
no
.
If the software of the FlowCheck sensor connection line/
FlowCheck sensor connection line with MEDUtrigger is not
up-to-date, the device updates the software before the key
function check starts.