4 Preparation and operation
2. Connect the open end of the nebulizer
,
or
with the
(breathing system filter, bacteria filter or a combined
breathing system/bacteria filter).
3. Place the filter
(breathing system filter, bacteria filter or a
combined breathing system/bacteria filter) on the breathing
circuit's FlowCheck sensor
4. When using the pneumatic medication nebulizer WM 15827:
Connect the flexible oxygen tube to the inhalation outlet on
the pressure reducer. When doing so, note: The nebulizer must
be in a horizontal position for sufficient nebulization to occur.
Result
A nebulizer is connected.
4.5 Switching the device on
Requirement
•
The device is disconnected from the patient.
•
A fully charged battery is inserted in the device.
•
The device is connected to the oxygen supply.
1. Briefly press the On/Off button
.
An automatic self-test starts, which runs through the following
sequence:
•
The alarm light flashes twice and two short test tones
are emitted
•
The start screen appears