Page 19
4.3
Installation preparations
1. Check and verify that the counter carrying the autoclave is a rigid and
leveled surface and can carry a load of 225 lb (102kg).
Attention:
The Elara11 is not designed for use on any standard slide out shelf. If it is
necessary to use a slide out shelf, it must be tested and/or rated for 225
Lbs or more.
2.
Check and verify that the counter dimensions are, at least, 22” wide x
24” deep (55cm x 63.5cm).
3. Keep the back and the sides of the autocla
ve approximately 4” (10 cm)
away from the wall to allow ventilation and facilitate the device
disconnection.
4. If placed in a cabinet, verify that the rear of the cabinet is open to allow
ventilation.
Note:
Insufficient space for ventilation may result in an increase of the autoclave's
temperature that may cause a malfunction or damage the instrument.
5. It is recommended that enough space be left around the autoclave to
give a technician access for servicing the machine.
6. Check and verify that the room ventilation is 10 cycles per hour
minimum.
7. Check and verify that the ambient temperature range is 41°F - 104°F
(5
º
C-40
º
C), it is preferable not to exceed 86°F (30
º
C).
8. Check and verify that the ambient relative humidity does not exceed
85%
9. To update the atmospheric pressure in the autoclave, keep the
chamber's door open for approx. 5 minutes. The system will be updated
automatically.
4.4
Placing the Autoclave
Keep all the sides of the autoclave approximately 3” (75.2 mm) away from the
wall to allow ventilation.
If placed in a cabinet, verify that the rear of the cabinet is open to allow
ventilation.
Insufficient space for ventilation may result in an increase of the autoclave's
temperature that may damage the instrument.
It is recommended that enough space be left around the autoclave to give a
technician access for servicing the machine.
4.5
Connections to Utility Supplies
1. Check and verify that the power supply is a 1 phase, 230Vac ±10%,
50/60Hz (as appropriate), 15A supply.
2. Check and verify grounding of the autoclave.
3. Check and verify that the electrical net is protected with a current
leakage safety relay.
www.shanghaidelan.com 021-54263653
Summary of Contents for ELARA11-D
Page 163: ...Page 162 Piping Diagram w w w s h a n g h a i d e l a n c o m 0 2 1 5 4 2 6 3 6 5 3...
Page 164: ...Page 163 w w w s h a n g h a i d e l a n c o m 0 2 1 5 4 2 6 3 6 5 3...
Page 165: ...Page 164 w w w s h a n g h a i d e l a n c o m 0 2 1 5 4 2 6 3 6 5 3...
Page 166: ...Page 165 w w w s h a n g h a i d e l a n c o m 0 2 1 5 4 2 6 3 6 5 3...
Page 167: ...Page 166 w w w s h a n g h a i d e l a n c o m 0 2 1 5 4 2 6 3 6 5 3...
Page 168: ...Page 167 w w w s h a n g h a i d e l a n c o m 0 2 1 5 4 2 6 3 6 5 3...
Page 169: ...Page 168 Electric Diagram w w w s h a n g h a i d e l a n c o m 0 2 1 5 4 2 6 3 6 5 3...
Page 170: ...Page 169 w w w s h a n g h a i d e l a n c o m 0 2 1 5 4 2 6 3 6 5 3...
Page 171: ...Page 170 w w w s h a n g h a i d e l a n c o m 0 2 1 5 4 2 6 3 6 5 3...
Page 172: ...Page 171 w w w s h a n g h a i d e l a n c o m 0 2 1 5 4 2 6 3 6 5 3...
Page 173: ...Page 172 w w w s h a n g h a i d e l a n c o m 0 2 1 5 4 2 6 3 6 5 3...
Page 174: ...Page 173 w w w s h a n g h a i d e l a n c o m 0 2 1 5 4 2 6 3 6 5 3...
Page 175: ...Page 174 w w w s h a n g h a i d e l a n c o m 0 2 1 5 4 2 6 3 6 5 3...
Page 176: ...Page 175 w w w s h a n g h a i d e l a n c o m 0 2 1 5 4 2 6 3 6 5 3...