Page 125
8. Install a new transducer.
9. Assemble the cap to the new transducer and tighten it with screw (1).
10. Reassemble the outer cover.
10.35
Replacing the pressostat
Caution!
Before starting, disconnect the instrument from the power source and
ensure that there is no pressure in the autoclave.
Allow the autoclave to cool before removing outer covers.
The pressostat’s function is to turn off power when the pressure is too high.
1. Open the outer covers of the autoclave.
2. Disconnect the wires (1).
3. Unscrew the nut (2).
Note:
Follow the steps in this section in the reverse order to install the
new pressostat.
4. Perform any cycle to verify that the pressostat operates as required.
1
2
3
4
5
6
www.shanghaidelan.com 021-54263653
Summary of Contents for ELARA11-D
Page 163: ...Page 162 Piping Diagram w w w s h a n g h a i d e l a n c o m 0 2 1 5 4 2 6 3 6 5 3...
Page 164: ...Page 163 w w w s h a n g h a i d e l a n c o m 0 2 1 5 4 2 6 3 6 5 3...
Page 165: ...Page 164 w w w s h a n g h a i d e l a n c o m 0 2 1 5 4 2 6 3 6 5 3...
Page 166: ...Page 165 w w w s h a n g h a i d e l a n c o m 0 2 1 5 4 2 6 3 6 5 3...
Page 167: ...Page 166 w w w s h a n g h a i d e l a n c o m 0 2 1 5 4 2 6 3 6 5 3...
Page 168: ...Page 167 w w w s h a n g h a i d e l a n c o m 0 2 1 5 4 2 6 3 6 5 3...
Page 169: ...Page 168 Electric Diagram w w w s h a n g h a i d e l a n c o m 0 2 1 5 4 2 6 3 6 5 3...
Page 170: ...Page 169 w w w s h a n g h a i d e l a n c o m 0 2 1 5 4 2 6 3 6 5 3...
Page 171: ...Page 170 w w w s h a n g h a i d e l a n c o m 0 2 1 5 4 2 6 3 6 5 3...
Page 172: ...Page 171 w w w s h a n g h a i d e l a n c o m 0 2 1 5 4 2 6 3 6 5 3...
Page 173: ...Page 172 w w w s h a n g h a i d e l a n c o m 0 2 1 5 4 2 6 3 6 5 3...
Page 174: ...Page 173 w w w s h a n g h a i d e l a n c o m 0 2 1 5 4 2 6 3 6 5 3...
Page 175: ...Page 174 w w w s h a n g h a i d e l a n c o m 0 2 1 5 4 2 6 3 6 5 3...
Page 176: ...Page 175 w w w s h a n g h a i d e l a n c o m 0 2 1 5 4 2 6 3 6 5 3...