Overview of operating table column

26

4990113_030_01 — 2069106 — 2017–11–02

3.2

Fixed installation operating table column with internal power supply

Operating table column is permanently installed in the operating 

room and permanently connected to the power supply of the room. 

The operating table column can be rotated continuously. The 

battery in the operating table column serves as an emergency 

power supply in the event of a power failure.

[15] Cover cap
[16] Ethernet connection socket
[17] Connection socket hardware signals

Key:

Key:

[1]

Column keypad

[4]

Connection socket control unit (on both sides of the column 

head)

[5]

Endolight

[6]

Mount wedge for operating tabletop

[7]

Floor plate

[8]

Column head

[9]

Lowest column cover (column foot)

[10] Electrical contact plate for the operating tabletop
[14] Reset button
[15] Cover cap
[18] Foot pedal brake

Summary of Contents for 1717020

Page 1: ...TruSystem 7500 Operating Table Column Instruction manual...

Page 2: ...I medical device according to the Council Directive 93 42 EEC concerning medical devices and is compliant with the Directive version currently in force at the time of product sale including its chang...

Page 3: ...nary column TruSystem 7500 1717020 Stationary column TruSystem 7500 U 1730731 TruSystem 7500 Hybrid SC 1854086 TruSystem 7500 Hybrid Plus SC 1854088 TruSystem 7500 Hybrid MR SC 2064886 TruSystem 7500...

Page 4: ...4 4990113_030_01 2069106 2017 11 02 This page is intentionally empty...

Page 5: ...ower supply 26 3 3 Fixed installation operating table column without internal power supply 27 3 4 Labels on the product 28 3 5 Audible signals factory setting 34 4 Control units 35 4 1 Malfunction cau...

Page 6: ...umn 59 12 8 Endolight 60 12 9 Localize wireless TruSystem 7500 remote control 60 12 10 Emergency mode 60 12 11 Further functionalities 62 13 Configuration options 63 14 Transfer shuttle operation 64 1...

Page 7: ...Mobile operating table 82 19 2 Fixed installation operating table 82 19 2 1 Operating table column with internal power supply 82 19 2 2 Operating table column without an internal power supply 82 20 D...

Page 8: ...Content 8 4990113_030_01 2069106 2017 11 02 This page is intentionally empty...

Page 9: ...y close attention to the information contained in this manual to reduce risk Information on the installation of a fixed installation operating table column can be found in the installation instruction...

Page 10: ...rt of the operating table Mobile operating table The operating table can be transported by a shuttle Fixed installation operating table The operating table column is permanently installed in the opera...

Page 11: ...the instruction manuals of the specific products VAH Verbund f r Angewandte Hygiene e V German Registered Association for Applied Hygiene Accessories accessory components Object designed by the manufa...

Page 12: ...or damage to property or injury to persons resulting from improper use 1 2 Contraindications Do not transport objects devices or materials on the operating table The operating table column is not appr...

Page 13: ...ction Improper loading may damage the operating tabletop or cause the mobile operating table to tip over When the tabletop has been moved in a longitudinal or transverse direction the patient may not...

Page 14: ...ompatibility Shuttle 3 6 360 1489491 Shuttle 2 7 1254362 Shuttle 2 7 360 1459862 Shuttle 3 7 1254363 Shuttle 3 7 360 1459863 Power Shuttle H 1774603 Designation Material number Designation see label o...

Page 15: ...4 1 Safety during operation These operating instructions must be read and understood in full by the operating personnel especially the chapter Safety instructions prior to start up or if appropriate...

Page 16: ...egible Replace damaged or loose labels Labels may not be changed removed or covered with other objects For the proper and safe use of the operating tabletop and additional equipment follow the instruc...

Page 17: ...vents the risk of electric shock Get a qualified electrician to check the electrical safety of the operating table and the power supply every year We recommend an annual general safety inspection by T...

Page 18: ...nel 1 4 3 High frequency HF surgical equipment and defibrillators The operating table is suitable for the use of HF surgical equipment defibrillators and defibrillator monitors Please follow the manuf...

Page 19: ...Medical service technicians or by personnel authorized and trained by Trumpf Medical Trumpf Medical will not be held liable for damage of any kind arising from the failure to perform inspections or a...

Page 20: ...pter 20 on page 84 1 5 Explanation of Symbols Important information in this instruction manual is marked with symbols and keywords Keywords such as DANGER WARNING or CAUTION indicate the level of dang...

Page 21: ...l applicable safety and maintenance instructions are strictly adhered to Trumpf Medical products will have a service life of 10 years The service life includes a warranty for product functionality in...

Page 22: ...uttle Clean the operating table column before using it for the first time see chapter 15 on page 73 mobile operating table column The delivery of the mobile operating table includes the operating tabl...

Page 23: ...st 12 hours 11 Switch on the operating table column at the column keypad Press the ON key 12 Press and hold the center pedal 41 at the shuttle Move the operating table column to the most elevated posi...

Page 24: ...20 Release the brake and set the required driving mode on the shuttle Note the operating instructions of the shuttle 21 Drive the shuttle with the operating table to the intended location of use 22 En...

Page 25: ...the operating table column allow use of the operating table without an external power supply Key 1 Column keypad 2 Connector socket for mains power cable 3 Connector pin for equipotential bonding cab...

Page 26: ...ed continuously The battery in the operating table column serves as an emergency power supply in the event of a power failure 15 Cover cap 16 Ethernet connection socket 17 Connection socket hardware s...

Page 27: ...pply unit with integrated batteries is located outside the operating room e g installed in the control cabinet The batteries in the power supply unit serve as an emergency power supply in case of powe...

Page 28: ...n Item Label Meaning Position mobile operating table column Fixed installation operating table column 20 Device label on the lowest column cover 21 Follow the instruction manual over the device label...

Page 29: ...variants Hybrid MR and Hybrid MR IMRIS 29 Transport of the operating table column is only permitted with the associated shuttle on the lowest column cover only U types of the operating table column 3...

Page 30: ...er 1 4 4 on page 16 over the device label on the lowest column cover only for operating table column TruSystem 7500 Hybrid MR external power supply unit 35 Device label on the housing 36 Follow the op...

Page 31: ...rmittent load 2 min ON 8 min PAUSE G Infrared code for the IR remote control corresponds to the last three digits of the serial number H The device is declared compliant in accordance with the Council...

Page 32: ...d stop time 2 min on 8 min off G Infrared code for the IR remote control corresponds to the last three digits of the serial number H Unique Device Identification UDI comprising Data Matrix Code 01 Glo...

Page 33: ...er 700 VA F Operating mode Continuous operation with intermittent load 2 min ON 8 min PAUSE G Unique Device Identification UDI comprising Data Matrix Code 01 Global Trade Item Number GTIN 11 Date of m...

Page 34: ...cending tone sequence Switch off the operating table Descending tone sequence End position or level position of selected adjustment range reached Single tone Confirmation of an operational procedure e...

Page 35: ...ion caused by other devices Portable and mobile RF communication devices may interfere with the medical device Examples of potential sources of interference Electronic control gear ECG for fluorescent...

Page 36: ...ating table press and hold the function key on the keypad until the desired position is reached The function stops when the key is released The keypad is locked again automatically ten seconds after t...

Page 37: ...per back section hinges up i19 Upper back section hinges down i20 Activate emergency mode display Emergency mode i21 Brake only for fixed installation operating table column Display Brake i44 Localize...

Page 38: ...the operating table using the column keypad is not possible Except for OFF key The operating table can be turned off at any time using the OFF key on the column keypad Press button i53 The keypad is r...

Page 39: ...d green Brake on the fixed installation operating table column is activated locked It is not possible to turn the column Flashing green Brake on the fixed installation operating table column is deacti...

Page 40: ...luminated No external supply voltage available The operating table is running on battery Charge mobile operating table column if required Restore the external power supply e g on off switch on the ope...

Page 41: ...12 Raise leg section hinges i13 Lower leg section hinges i14 Lower back section hinges up i15 Lower back section hinges down i16 Longitudinal travel toward head end i17 Longitudinal travel toward foot...

Page 42: ...while using the operating table 5 1 2 Fixed installation operating table column Two lithium ion batteries provide the internal power supply for the fixed installation operating table column With the s...

Page 43: ...erating table otherwise in case of a fault the permitted patient leakage current threshold for cardiac surgery applied part CF according to IEC 60601 1 will be exceeded The fault occurs when the prote...

Page 44: ...nd an audible signal sounds 2 Pull the mains power cable plug from the connector socket on the column foot 5 2 2 Fixed installation operating table The fixed installation operating table column is per...

Page 45: ...nds that repeats at intervals of a few minutes 1 Set the external power supply connection for the operating table see page 43 Indicator a3 on the column keypad lights up Indicator a1 on the column key...

Page 46: ...g 12 to the equipotential connection pin 3 of the operating table column Plug the other end of the cable into the equipotential bonding in the room The connection location on the column is identified...

Page 47: ...r It may be necessary for an operating table in battery mode to be switched on again in the interim for longer surgeries without the operating table controls being used 8 2 Fixed installation operatin...

Page 48: ...must be nearly 100 available Interruptions to the network may only take a few minutes per year Only one system agent or PEMS manufacturer may be responsible for the network The operating table and med...

Page 49: ...he display of the TruSystem 7500 remote control The displays are not available with the TruSystem 5500 remote control Displays on the remote control The external control of the operating table can lea...

Page 50: ...has been activated Note the instructions for use of the connected device The operation of the operating table with column keypad and transfer pedal at the shuttle is possible in emergency situations...

Page 51: ...ng table Take appropriate precautions for the safety of personnel Note the instructions for use of the connected device The operation of the operating table with column keypad and transfer pedal at th...

Page 52: ...he cover cap 15 made of plastic and press the button 14 briefly using a pointed object e g a ballpoint pen Then switch the operating table on again With an external power supply the operating table sw...

Page 53: ...olumn keypad can only be selected once the keys are unlocked The locked keypad functions can also be selected using two key control Press the function key on a control unit until the table is in the r...

Page 54: ...the column keypad on an operating table column without the operating tabletop 12 1 Key release The column keypad is locked by default so that the functions on the operating table cannot be accidental...

Page 55: ...i4 on the column keypad does not correspond to the operating table head position of the patient Some operating table functions e g Trendelenburg tilting are then laterally inversed In such cases you...

Page 56: ...operating table during positioning 3 Press key i11 You will hear an audible signal when the final level position has been reached Indicator i11 on the column keypad lights up A warning tone sounds hi...

Page 57: ...op 12 5 Anti Trendelenburg Trendelenburg The Trendelenburg position tilts the operating tabletop about the transverse axis toward the head end Reverse Trendelenburg tilts toward the foot end Setting t...

Page 58: ...ction i5 Lift i6 Lower Press key i5 or i6 CAUTION Hazard to the patient due to risk of collision Risk of entrapment for personnel Lowering the operating tabletop can result in collisions with the floo...

Page 59: ...ed The display i21 on the column keypad and the status bar on the TruSystem 7500 remote control unit is flashing 2 Manually rotate the operating table column into the desired position 3 Press key i21...

Page 60: ...here is a defect in the central operating table control system emergency operation makes it possible to complete the ongoing surgery and remove the patient Restriction In the emergency mode the operat...

Page 61: ...In emergency mode the patient is transferred as follows 1 Return operating table to level position If necessary adjust the functions individually 2 For shuttle types with tilt and height adjustment s...

Page 62: ...le column 62 4990113_030_01 2069106 2017 11 02 12 11 Further functionalities The possible functions that can be used on the operating table depend on the operating tabletop used and are described in t...

Page 63: ...t individually and not at the level position A message appears on the TruSystem 7500 remote control display For additional adjustment beyond the threshold value briefly release the function key and pr...

Page 64: ...ut the terminal box As soon as the shuttle is at the operating table column all functions on the operating table are blocked and only the transfer can be carried out WARNING Risk of injury during the...

Page 65: ...ansfer to the end position and stop the function before a dangerous situation arises In an emergency release the pedal or the key The process stops Pressing the middle pedal or the key i11 level posit...

Page 66: ...vy patients pay attention to the central position of the patient center of gravity above the mount wedges Otherwise no transfer can take place 14 1 1 Transfer from the shuttle to the operating table c...

Page 67: ...matically In exceptional cases the sensors of the operating table column cannot recognize the tabletop position The message Table Top Orientation undefined please select manually appears on the TruSys...

Page 68: ...41 down at the shuttle until the transfer is complete The transfer can also be made with the key i11 level position on the remote control The operating table column moves to the level position and the...

Page 69: ...ing table column when the section segments are inclined downwards If necessary place the section segments on the operating tabletop horizontally and move the operating table column upwards 3 On the sh...

Page 70: ...rating table 14 2 2 Set down operating table 1 Clear the area for the operating table The operating table may not be placed on cables or objects Set down operating table Press and hold the center peda...

Page 71: ...een integrated into the operating table control After the transfer has been activated the leveling stops in case of a defective operating table drive Release the center pedal at the shuttle or with th...

Page 72: ...ck for collision appears on the TruSystem 7500 remote control display The emergency transfer is carried out half speed and a continuous shrill warning sound can be heard The emergency transfer is term...

Page 73: ...damp cloth In this CAUTION Risk of physical injury Disconnect the operating table from the external power supply before performing any maintenance work or cleaning Ensure that no motorized movements a...

Page 74: ...and disinfecting remove any excess moisture such as drops at the bottom edge using a dry cloth Remove the pad to clean the pad plate and the underside of the pad Attach the cleaned and disinfected pad...

Page 75: ...table in the form of pictograms 16 2 Conditions for operation Pictogram Meaning Temperature range for storage and transport 15 C to 55 C 5 F to 131 F Air humidity for storage and transport 5 to 95 Air...

Page 76: ...on the operating table equipment and patient weight Fixed installation operating table column Date of manufacture See device label on the product Column diameter length x width 372 mm x 311 mm Diamete...

Page 77: ...the longitudinal axis 30 to the left 30 to the right Rotation of the fixed installation operating table column SF SF U Freely rotatable 360 All variants SB and Hybrid 350 Classification Operating tab...

Page 78: ...ical and electromagnetic interference levels or may lead to the reduced interference immunity of the operating table Repairs on remote controls especially on the cables may only be performed by the Te...

Page 79: ...10 AT is defective Contact Technical Customer Service Operating table cannot be run on line power EXTERNAL POWER SUPPLY indicator does not illuminate The mains power supply cable is defective Contact...

Page 80: ...hen leg sections move together Possible risk of collision signal is warning OK not an error Transfer of the operating tabletop from the operating table column to a shuttle is not possible Drive defect...

Page 81: ...ions to the product Please contact the Technical Customer Service at Trumpf Medical if you require service The following maintenance intervals have been established for the operating table First maint...

Page 82: ...2 Press key i2 on the column keypad or remote control for more than 2 seconds An audible signal sounds shortly before shutdown 3 Charge the operating table every 3 months Establish an external power s...

Page 83: ...unit every 3 months see page 45 Establish an external power supply connection for the power supply unit on off switch e g on the operating panel Follow the charging cycle If a charging cycle is longe...

Page 84: ...onal authority Trumpf Medical will take back your old equipment or products that are defective or no longer used Please contact Technical Service in Saalfeld for more information When the operating ta...

Page 85: ...as specified below The customer or user of the aforementioned device should ensure that it is operated in one of the environments as described Emitted interference measurements Compliance Electromagn...

Page 86: ...ages surges in accordance with IEC 61000 4 5 1 kV voltage outer conductor outer conductor 2 kV voltage outer conductor earth 1 kV voltage outer conductor outer conductor 2 kV voltage outer conductor e...

Page 87: ...data and d is the recommended safe distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each f...

Page 88: ...are controlled The customer or user of the above mentioned device can contribute towards avoiding electromagnetic interference by observing the minimum distance between portable and mobile HF telecomm...

Page 89: ...This page is intentionally empty...

Page 90: ...es and Branch Offices France TRUMPF Syst mes M dicaux United Kingdom TRUMPF Medical Systems Ltd Italy TRUMPF MED ITALIA s r l United States of America TRUMPF Medical Systems Inc Australia TRUMPF Med A...

Reviews: