Important information
18
4990113_030_01 — 2069106 — 2017–11–02
• Ensure that the power cable and equipotential bonding cable
are not crushed or driven over. Stop using cables if damaged. If
you have any concern about the safety of the power cables and
equipotential bonding conductors, work only with the internal
(rechargeable) power supply until you have replaced the cable
in question.
Remove cables prior to relocating the table.
• Always keep the operating table column in an operational
(charged) state so that it is ready for use immediately! Fully
charge the operating table column prior to first use.
Fixed installation
operating table
column
• In an emergency, disconnect the operating table column from
the power supply using the separate on/off switch (e.g. on the
operating panel).
1.4.3
High-frequency (HF) surgical equipment and defibrillators
The operating table is suitable for the use of HF surgical
equipment, defibrillators, and defibrillator monitors. Please
follow the manufacturer's instruction manual and safety
instructions for the equipment.
Trumpf Medical operating tables are electrically conductive in
accordance with the current version of the IEC 60601-1
regulations. When high-frequency surgery devices, defibrillators,
and defibrillator monitors are used, the patient runs the risk of burns
if safety precautions are not followed.
• Risk of burns to patient! Position the patient on the operating
table so that he or she is insulated from metal parts (operating
table, accessories) and conductive pads or hoses.
• Make sure the patient does not come into contact with damp
towels or pads! Only use dry materials!
• The electrically motorized operating table functions can be
interrupted when high-frequency surgical devices are used
simultaneously.