Operate fixed installation operating table column
54
4990113_030_01 — 2069106 — 2017–11–02
The speed at which the function is executed depends on the
patient's weight. The speed may be lower for a heavier patient.
The possible functions that can be used on the operating table
depend on the operating tabletop used and are described in the
operating tabletop instruction manual. Only the level position and
lift functions can be selected via the column keypad on an
operating table column without the operating tabletop.
12.1
Key release
The column keypad is locked by default so that the functions on the
operating table cannot be accidentally activated. It is not possible
to control the operating table via the column keypad in this
condition. Indicator [i53] does not light up. The operating table can
be switched off at any time using the
O
FF
key on the column
keypad, regardless of the key release.
Releasing the key
lock
Press key [i53]. Indicator [i53] lights up. The keypad is released for
operation. The keypad automatically locks again 10 seconds after
the last button operation.