Important information
4990113_030_01 — 2069106 — 2017–11–02
17
Pay particular attention to the operating table when lowering,
since the operating tabletop may collide with objects lying
below it, or it may even make contact with the floor if set to an
extreme position. In some circumstances, there is a possible
crushing hazard for persons.
• When moving to the level position on the operating table, larger
patient incline positions can occur compared to the initial
position. Only select the function Level position (transfer position)
if collisions cannot occur. Monitor the movement to the level
position up to the final position.
Only in case of mobile operating table column:
• Danger of collisions when transporting the operating table. Be
careful to avoid collisions with nearby persons or furniture and to
prevent damage to the table!
1.4.2
Electrical safety
• Position the operating table such that the detachable plug or
device couplers are easily accessible.
• The operating table may only be connected to a power supply
with a protection conductor. This prevents the risk of electric
shock.
• Get a qualified electrician to check the electrical safety of the
operating table and the power supply every year.
We recommend an annual general safety inspection by
Trumpf Medical Technical Customer Services.
• Always keep the wireless remote control in an operational
(charged) state so that it is ready for use immediately!
mobile operating
table column
• When used on an electrically conductive floor, the operating
table may be used only if connected to an equipotential
bonding cable. The equipotential bonding cable connection
location on the operating table is in accordance with
IEC 60601-1 (DIN EN 60601-1/VDE 0750-1).
• An extension cable cannot be used with the mains power cable
at the operating table, otherwise, in case of a fault, the permitted
patient leakage current threshold for cardiac surgery (applied
part CF according to IEC 60601-1) will be exceeded. The fault
occurs when the protection conductor is interrupted by a
damaged mains power cable. The operating table must only be
connected to the power supply using the original mains power
cable from Trumpf Medical.
In addition, the supply voltage must be observed. The operating
table must only be connected to the supply voltage indicated on
the device label.