Power supply
44
4990113_030_01 — 2069106 — 2017–11–02
be outside of an area where there is a risk of explosion. There
is risk of explosion in conjunction with flammable mixtures of
anesthetics and air.
4. Indicators [i1] and [a3] on the column keypad light up.
After a few seconds, an audible signal sounds. The operating
table is only ready for use once the signal tone has sounded.
If the operating table was switched off, it will switch on
automatically when plugged into the wall outlet.
Disconnecting the
operating table from
the power supply
Pull the mains power cable plug from the grounded wall
socket. The display [a3] on the column keypad goes out and
an audible signal sounds.
2. Pull the mains power cable plug from the connector socket on
the column foot.
5.2.2
Fixed installation operating table
The fixed installation operating table column is permanently
connected to the external power supply of the room and is only
disconnected by means of the separate on/off switch (e.g. on the
operating panel) in the event of an emergency or for service
purposes. When connected to the power supply, there are no
limitations on the use of electrically powered table functions.
1.
WARNING
Risk of explosion due to flammable gas
mixtures!
There is a risk of explosion in conjunction with flammable
mixtures of anesthetics and air if the mains power cable is first
disconnected from the operating table and then from the
socket. The sequence for disconnecting the mains power
cable must be strictly followed.