Switch off the operating table
4990113_030_01 — 2069106 — 2017–11–02
47
8
Switch off the operating table
8.1
mobile operating table column
The operating table can be switched off only if it is running on
battery power. If connected to the external power supply, the
operating table remains on permanently, since it will turn itself back
on automatically after being turned off.
1. Disconnect the operating table from the mains (see page 44).
2. Press key [i2] on the column keypad or remote control for
more than 2 seconds. An audible signal sounds shortly before
shutdown.
In the case of an operating table column (as of software
version 6.x.x.x) with the operating tabletop placed on the table, the
operating table column in battery mode automatically switches off
for 8 hours after the last key press on a control unit to protect the
batteries. In the case of an operating table column without an
operating tabletop, the automatic switch-off takes place after
1 hour.
It may be necessary for an operating table in battery mode to be
switched on again in the interim for longer surgeries without the
operating table controls being used.
8.2
Fixed installation operating table column with internal power supply
The operating table column is permanently connected to the power
supply of the room and always remains switched on. After switching
off, the operating table column automatically switches back on.
8.3
Fixed installation operating table column without internal power supply
Press key [i2] on the column keypad or remote control for more than
2 seconds. An audible signal sounds shortly before shutdown.
In the case of an operating table column (as of software
version 6.x.x.x) with the operating tabletop placed on the table, the
operating table column in battery mode automatically switches off
for 8 hours after the last key press on a control unit to protect the
batteries. In the case of an operating table column without an
operating tabletop, the automatic switch-off takes place after
1 hour.
The external power supply unit is permanently connected to the
power supply of the room and always remains switched on.