Control units
4990113_030_01 — 2069106 — 2017–11–02
39
Level position [i11]
Illuminated green The operating table is in level
position.
–
Flashing
green Leveling of the operating table is
incomplete.
Perform an emergency
transfer (see page 71) if
leveling is incorrect.
Emergency mode [i20]
Illuminated red
In addition to the
E
MERGENCY
MODE
display, the
READY
TO
OPERATE
[i1] display flashes, while
the
MALFUNCTION
[a4] display
lights up.
The operating table emergency
mode was activated manually.
Block the defective operating
table from use for subsequent
operations.
Contact Technical Service.
Brake [i21] (only active for fixed installation operating table column)
Illuminated green Brake on the fixed installation
operating table column is
activated (locked). It is not
possible to turn the column.
–
Flashing
green Brake on the fixed installation
operating table column is
deactivated (released). It is
possible to turn the column.
–
Battery status [a1]
The display is only available for the operating table column with internal power supply (battery in the
column).
Illuminated green Operating table column is fully
charged.
–
Flashing
green Operating table column is
charging.
–
Illuminated red
Charging of the operating table
column is required.
Establish external power
supply connection for
operating table.
Flashing
red
Operating table column
battery is depleted! Electrical
functions are severely limited.
Operating table shutdown is
imminent.
Establish external power
supply connection for
operating table.
Indicator
Status
Color Meaning
Action