Power supply
4990113_030_01 — 2069106 — 2017–11–02
43
5.2
External power supply (line connection)
5.2.1
Mobile operating table column
The external power supply of the operating table column is
provided by the power supply in the room. When connected to the
power supply, there are no limitations on the use of electrically
powered table functions.
Connecting the
operating table to
the power supply
Flip up the connector socket [13] lid in the column foot [9] and
insert the mains power cable plug [11] into the bushing as
shown [2]. The connection location is identified by the symbol
for the power supply connection.
3. Route the cable such that no one can trip or fall over it.
Plug the connector of the mains power cable into a grounded
power socket in the room. The grounded power socket must
CAUTION
Hazard to the patient!
An extension cable cannot be used with the mains power cable at
the operating table, otherwise, in case of a fault, the permitted
patient leakage current threshold for cardiac surgery (applied part
CF according to IEC 60601-1) will be exceeded. The fault occurs
when the protection conductor is interrupted by a damaged mains
power cable. The operating table must only be connected to the
power supply using the original mains power cable from
Trumpf Medical.
In addition, the supply voltage must be observed. The operating
table must only be connected to the supply voltage indicated on
the device label.
1.
DANGER
Risk of death from damaged mains power
cables!
There is a risk of death by electric shock if the mains power
cable is damaged. Check the mains power cable and
remove from use if it has been crushed or if the insulation is
damaged.
2.
WARNING
Risk of explosion due to flammable gas
mixtures!
There is risk of explosion in conjunction with flammable
mixtures of anesthetics and air if the mains power cable is first
plugged into the socket and then into the operating table. The
sequence for connecting the mains power cable must be
strictly followed.