Equipotential bonding of the mobile operating table column
46
4990113_030_01 — 2069106 — 2017–11–02
6
Equipotential bonding of the mobile operating table column
Equipotential bonding eliminates the potential differences between
the touchable conducting parts near the patient if the operating
table is connected to the equipotential bonding in the room.
Connect the equipotential bonding cable plug [12] to the
equipotential connection (pin) [3] of the operating table column.
Plug the other end of the cable into the equipotential bonding in the
room. The connection location on the column is identified by the
symbol for the equipotential bonding cable connection.
7
Switch on the operating table
Press the power key [i1] on the column keypad. Indicator [i1] on the
column keypad lights up and an audible signal sounds after a few
seconds. The operating table is only ready for use once the signal
tone has sounded. The indicator [i1] lights up continuously when the
operating table column is switched on.