Operate fixed installation operating table column
4990113_030_01 — 2069106 — 2017–11–02
61
Activating the
emergency mode
The emergency mode has to be manually activated by the user:
1. Unlock the column keypad. Press key [i53] for this purpose.
2. Press key [i20] on the column keypad for at least 10 seconds.
A warning tone sounds and constantly repeats.
The emergency mode is confirmed by a double tone and is
then activated. The displays [i20] and [a4] light up
continuously and the display [i1] flashes.
The operating table can be restored to its normal mode of
operation by switching it off and then on again.
Transfer in
emergency mode
In emergency mode, the patient is transferred as follows:
1. Return operating table to level position. If necessary, adjust the
functions individually.
2. For shuttle types with tilt and height adjustment, set the level
position on the shuttle.
Transfer is only possible if the mount wedges on the shuttle are
horizontal and the shuttle column is in the uppermost position.
3. Carefully move the shuttle up to the position stop of the
operating table column. Caution: there is no monitoring in the
emergency mode as to whether the shuttle is correctly
positioned on the operating table column.
Press the key [i6] (lift down) on the column keypad. The
operating tabletop is transferred to the shuttle.
4.
CAUTION
Risk of crushing injury to hands!
Do not get your hands caught between the operating table
and the shuttle during the transfer.