Technical data
4990113_030_01 — 2069106 — 2017–11–02
77
Radio transmission,
operating table
column
Transmission
frequency band
2.405 GHz to 2.480 GHz
Modulation type O-QPSK (Offset Quadrature Phase Shift Keying)
HF bandwidth
2 MHz (IEEE 802.15.4)
Output power
(Watt EIRP)
0.000603
Emission
designation
1M67G1D
Setting ranges
Lift
550 mm
Reverse Trendelenburg/Trendelenburg
(inclination about the transverse axis)
45°/45°
Tile (inclination about the longitudinal
axis)
30° to the left / 30° to the right
Rotation of the fixed installation
operating table column
SF, SF U
Freely rotatable
360°
All variants, SB and
Hybrid
350°
Classification
Operating table protection
classification
I
Device with internal power supply I.P.S (not for
operating table column TruSystem 7500 Hybrid
MR)
Level of protection against electrical
shock for the entire operating table
Applied part type B
Patient leakage current in accordance with CF in
compliance with IEC 60601-1
Degree of protection from water
penetration
IP X4
Operating mode
Continuous operation with intermittent load
2 min ON, 8 min OFF