Reset
52
4990113_030_01 — 2069106 — 2017–11–02
10
Reset
If the operating table reacts unexpectedly to a control, switch off the
table and switch it back on. If you cannot even switch it off, press
the reset button. The reset key [14] is located on the lowest column
cover [9] on the operating table side to the right of the column
keypad [1]. Remove the cover cap [15] (made of plastic) and press
the button [14] briefly using a pointed object (e.g., a ballpoint pen).
Then switch the operating table on again. With an external power
supply, the operating table switches on again automatically.
11
Load capacity
Use of the operating table system is permitted for a maximum
patient weight of 400 kg. Patient weight can be extremely restricted
by the operating tabletop. Note operating instructions for the
operating tabletop. The terms and conditions for equipping the
operating tabletop must be observed.