Transfer (shuttle operation)
4990113_030_01 — 2069106 — 2017–11–02
67
operating table column. The message
▸
Table top shuttle
present, transfer mode active
◂
appears on the
TruSystem 7500 remote control display.
5. Carry out the transfer: Press and hold the center pedal [41]
down at the shuttle until the transfer is complete. The transfer
can also be made with the key [i11] (level position) on the
remote control.
The operating table column moves into the upper lift position
and the operating tabletop is taken over by the column. Ensure
that the operating tabletop is correctly taken over from the
mount wedges of the operating table column. An acoustic
signal indicates the end of the transfer and the message
▸
OR
Table Top (type) recognized.
◂
appears on the
TruSystem 7500 remote control display.
The direction of the operating tabletop (head position) is
detected automatically. In exceptional cases, the sensors of
the operating table column cannot recognize the tabletop
position. The message
▸
Table Top Orientation undefined,
please select manually
◂
appears on the TruSystem 7500
remote control display. The display [a4] and both head
position displays [i3]/[i4] on the column keypad light up
alternatingly. In this case, you must manually switch the head
position according to the position of the patient's head. Press
key [i3] or [i4].
6. Drive the shuttle away.
14.1.2
Transfer of the operating table column to the shuttle
1. For shuttle types with tilt and height adjustment, set the level
position on the shuttle.
Transfer is only possible if the mount wedges on the shuttle are
horizontal and the shuttle column is in the uppermost position.
2. Caution: the shuttle can bump into the operating tabletop and
cannot be fully moved to the operating table column when the
section segments are inclined downwards. If necessary, place
the section segments on the operating tabletop horizontally
and move the operating table column upwards.