Operate fixed installation operating table column
56
4990113_030_01 — 2069106 — 2017–11–02
– The longitudinal slide and transverse slide of the operating
tabletop are brought to the starting position (depending on the
equipment of the operating table). If the extension adapter is
attached to the operating tabletop, the longitudinal travel is not
zeroed.
The functions are partially addressed simultaneously or
successively.
For a completely level position on the operating table, the
mechanical hinges on the operating tabletop must be brought
manually to level position.
Setting the function
1. Adjust the spreadable leg sections so that they are not on top
of one another.
2. Arrange accessories on the operating table in such a manner
that they do not collide with other accessories, section
segments or the operating table during positioning.
3. Press key [i11]. You will hear an audible signal when the final
level position has been reached. Indicator [i11] on the column
keypad lights up.
A warning tone sounds (high-pitched single tone) during the
movement, as soon as the leg section hinges are approaching
one another on the operating tabletop.
4. The section segments on mechanically adjustable hinges
(depending on the equipment of the operating table) must be
brought into the level position manually (horizontal position).
The line marking on the gearing corresponds to the level
position.
5. Move the longitudinal displacement manually into the level
position with a mechanical operating tabletop. See operating
instructions for the operating tabletop.
CAUTION
Risk to the patient due to the tilted position!
When approaching the level position, patient positions may
become more skewed than they were in the initial position. Monitor
the patient's position and stop the function before a dangerous
situation arises.