Special features of the operating table column
4990113_030_01 — 2069106 — 2017–11–02
49
9.2
External control
In the hybrid operating room, the operating table is connected to
another medical device (e.g. imaging system) via a network. The
medical device can block the operation of the operating table in
certain operating conditions by blocking functions. The pictogram
of the blocked function is no longer visible on the remote control of
the operating table and the function cannot be selected. In
emergency situations, the column keypad and the center pedal
(transfer) can still be used at the shuttle in all operating conditions of
the operating table.
The operating condition of the operating table can be switched
over as required by the connected medical device. The operating
table is thus determined by means of its own or third-party control
depending on the operating condition. The current operating
condition of the operating table is indicated in the display of the
TruSystem 7500 remote control. The displays are not available with
the TruSystem 5500 remote control.
Displays on the
remote control
The external control of the operating table can lead to unsafe
operating conditions for the operator and patient.
The unsafe operating conditions are displayed with a red display
background. The color of the background changes back to the
default as soon as the unsafe operating condition has ended.
CAUTION
Hazard to the patient due to improper operation!
Mobile operating table: Do not connect or disconnect the data
lines from the operating table during use.
CAUTION
Hazard to the patient due to improper operation!
Move the operating table with special care in emergency
situations. Instructions for use and the current operating condition of
the connected device.
Standard display
red display
Display
Meaning
Enter the following
code on the
connected system:
1234
The connected medical device would like
to exchange data with the operating
table. The displayed code must be
entered on the connected medical device
for safety reasons.
Note the instructions for use of the
connected device.
Color: Standard