Operate fixed installation operating table column
58
4990113_030_01 — 2069106 — 2017–11–02
For safety reasons, the combination of reverse/Trendelenburg and
transversal shift on the operating table is only possible to a limited
degree when using certain tabletops. Note operating instructions
for the operating tabletop.
12.6
Lift
The lift function raises or lowers the entire operating tabletop.
Setting the function
[i5] Lift
[i6] Lower
Press key [i5] or [i6].
CAUTION
Hazard to the patient due to risk of collision!
Risk of entrapment for personnel!
Lowering the operating tabletop can result in collisions with the
floor, furniture, or devices located beneath the operating tabletop.
Pay particular attention to leg sections that are lowered. Monitor
all movements on the operating table up to the end position and
stop the function before a dangerous situation arises. Clear the
area beneath the tabletop or raise the corresponding section
segments (leg/back section hinges). Next, continue moving the
operating tabletop downward.