OMNICxi Software Operation
Acquire and Analyze Data
Thermo Scientific
DXRxi User Guide
157
14. Click
COLLECT Cross Section
to collect the chemical image.
The image appears in the left pane. The cross sectional chemical image shows depth
(Y-, Z-axis) information rather than the normal surface (X-, Y-axis) information acquired
in
Image Regions mode
. Here is an example of a cross sectional image using the
. Each color in the chemical image represents a new layer in the sample.
Figure 91.
MCR cross sectional image of an optically transparent sample
15. Adjust the
setting if needed to refine the chemical image.
16. If you adjusted a setting, click
COLLECT Cross Section
again to recollect the image
data.
White (spectrum at
current cursor location)
Image Cross Section mode
MCR profile with 5
expected components
Red (calculated pure component
spectrum for red layer)
Cross sectional
chemical image
Current cursor
location
Summary of Contents for DXR xi
Page 12: ...1 Introducing the DXRxi Raman Imaging System Trademarks 6 DXRxi User Guide Thermo Scientific...
Page 14: ...2 DXRxi Microscope Operation 8 DXRxi User Guide Thermo Scientific...
Page 64: ...Safety Information 58 DXRxi User Guide Thermo Scientific Microscope Safety Labels or...
Page 65: ...Safety Information Thermo Scientific DXRxi User Guide 59...
Page 66: ...Safety Information 60 DXRxi User Guide Thermo Scientific...
Page 67: ...Safety Information Thermo Scientific DXRxi User Guide 61...
Page 68: ...Safety Information 62 DXRxi User Guide Thermo Scientific...
Page 166: ...3 OMNICxi Software Operation Acquire and Analyze Data 160 DXRxi User Guide Thermo Scientific...