Installation and Setup
Startup
2-8
146iQ Instruction Manual
Thermo Scientific
Use the following procedure when starting the instrument.
1.
Turn the power ON.
2.
Allow 90–120 minutes for the instrument to stabilize. During the time
that the instrument is warming up, the mode “warm up” is displayed
on the gas mode button in the title bar, and the concentration
calculation is turned off. To disable warm up, go to
Settings>
Configuration
.
3.
Allow 90 minutes for the instrument to stabilize (if the
ozonator/photometer options are installed).
4.
Set instrument parameters such as cylinder setup, gas setup, and
operation modes to their appropriate settings. For more information
about instrument parameters, see the “
Operation
” chapter.
Figure 2–7.
Front Panel and Touchscreen Display
Startup