Servicing
Optional Photometer DMC
7-38
146iQ Instruction Manual
Thermo Scientific
Use the following procedure to remove the pre-amplifier/detector assembly.
Equipment required:
Hex drive, 3/32
1.
Turn instrument OFF, unplug power cord, and remove the cover
(
Figure 2–1
).
2.
Unplug electrical cable from detector 1.
Figure 7–43.
Preamp Detector Cover Removal
Detector Board
Replacement