Optional Equipment
Connecting External Devices
9-4
146iQ Instruction Manual
Thermo Scientific
The Analog I/O Board consists of:
●
4 Isolated Analog Voltage Inputs, Input Voltage Range: 0–10 V
●
6 Isolated Analog Voltage Outputs, Three Ranges: 0–1.0 V, 0–5.0 V,
0–10 V
●
6 Isolated Analog Current Outputs, Two Ranges: 0mA–20mA, 4mA–
20mA
Table 9–3
lists the analog voltage inputs are used to monitor four external
0–10V signals.
Figure 9–4.
Analog Voltage Inputs
Table 9–3.
Analog Voltage Inputs Assignment
Terminal Number
Signal Name
1
Analog In 1
2 Analog
GND
3
Analog In 2
4 Analog
GND
5
Analog In 3
6 Analog
GND
7
Analog In 4
8 Analog
GND
Analog I/O Board
Analog Voltage Inputs