Servicing
Optional Zero Mix Valve Replacement
7-54
146iQ Instruction Manual
Thermo Scientific
Use the following procedure to replace the zero mix valve.
Equipment required:
Phillips drive, #2
Hex drive, 9/64
Adjustable wrenches, qty 2
1.
Turn instrument OFF, unplug power cord, and remove the cover
(
Figure 2–1
).
2.
Unplug zero mix valve. Refer to
Figure 7–57
and
Figure 7–58
.
3.
Using two adjustable wrenches (one to hold fitting body and one to
loosen the nut), disconnect plumbing 3X.
4.
Using #2 Phillips drive, unfasten two captive hardware.
Figure 7–61.
Replacing the Zero Mix Valve, Electrical
5.
Using a 9/64 hex drive, unfasten two 8-32 soc cap head screws.
6.
Pull zero mix valve upwards away from bracket.
Optional Zero
Mix Valve
Replacement