Operation
Settings
3-88
146iQ Instruction Manual
Thermo Scientific
The Permeation Oven Setup screen allows the user to define the
permeation tube gas, enter the permeation constant (K, described in
Chapter 9, “
Optional Equipment
”) and the relevant permeation rate of the
permeation tube, set the permeation oven temperature setpoint, and set up
the zero mode flows and up to 6 diluted span concentrations. The
respective minimum and maximum allowable values are calculated based
on setpoints and/or acceptable MFC operation ranges.
Use the
and
buttons to move up and down and the
and
buttons to move left and right.
Home Screen>Settings>Permeation Oven Setup
The Permeation Oven Setup screen contains the following information:
●
Across:
●
Permeation Oven:
This column lists items associated with the
permeation oven setup.
●
Setting:
User enters parameters relevant to the permeation oven
operation.
●
Minimum:
Minimum allowable value.
●
Maximum:
Maximum allowable value.
●
Units:
Displays units for each item.
●
Total Flow (sccm):
Displays the relevant measured/calculated total
flows based on flow measurement/chosen setpoint.
●
Down:
●
Permeation Gas Name:
User enters permeation tube gas name
through an alphanumeric keypad.
Permeation Oven
Setup
Summary of Contents for 146iQ
Page 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Page 2: ......
Page 8: ......
Page 22: ......
Page 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Page 160: ......
Page 170: ......
Page 176: ......
Page 252: ......
Page 300: ......
Page 301: ......