Operation
Calibration
3-12
146iQ Instruction Manual
Thermo Scientific
The Manual Ozonator Calibration screen allows the user to manually
calibrate ozonator drive power to O
3
concentration.
Use the
and
buttons to move up and down and the
and
buttons to move left and right.
Home Screen>Calibration>Manual Ozonator Calibration
Home Screen>Calibration>MFC Calibration>Continue to Ozonator
Lamp Power to Concentration Screen
The Ozonator Lamp Power to Concentration Screen contains the
following information:
●
Across:
●
Ozonator Drive %:
Output power of the ozonator.
●
Select Drive:
User initiates ozonator output power setting by
pressing Start for the corresponding ozonator drive %.
●
Enter O3 Concentration (ppb):
Allows user to enter calibrated
external photometer reading for the corresponding ozonator drive
% setting.