Calibration
Calibration of the individual MFC outputs
4-2
146iQ Instruction Manual
Thermo Scientific
3.
From the Home Screen, press Calibration>
MFC Calibration
to see the
screen as below:
4.
Choose the calibration temperature and pressure (this should reflect
what the flow meter is using, standard or volumetric).
5.
If using measured, enter the barometric pressure and ambient
temperature as measured by the flow meter or appropriate device under
the user defined temperature and pressure buttons after disabling the
‘Standard Temp and Pressure’ option, as shown below:
6.
Choose the MFC to be calibrated under the ‘Calibration’ column, as
described in the “
Operation
” Chapter, and continue to the MFC
calibration to see the screen below: