Introduction
Specifications
1-10
146iQ Instruction Manual
Thermo Scientific
Detection Limit
1.5 ppb (60 sec. averaging)
Precision 1
ppb
Zero Drift
<1.0 ppb (24 hours)
Span Drift
<1% of full scale (1 month)
Response Time
Photometer
System
20 seconds (10 sec. averaging)
<180 seconds
Linearity
±
1% full-scale
Flow Rate
0.8 LPM nominal
Stability and Repeatability
Meets or exceeds U.S. EPA requirements for a
Transfer Standard
Permeation Oven
Temperature Control
Three user selectable set points: 30, 35, 45
C
Temperature Stability
±
0.1
C
Warm-up Time
1 hour (permeation device can take 24 to 48
hours to stabilize)
Carrier Gas Flow
150 sccm nominal
Chamber size
Accepts permeation tubes up to 9 cm in total
length; 1 cm in diameter
Remote Operation
External Solenoid Outputs
Up to (8) 24 VDC external solenoids can be
controlled by 146iQ
Analog I/O
4 Isolated Voltage Inputs 0–10 V
6 Isolated Analog Voltages Outputs, with 4
selectable ranges
6 Isolated Analog Current Outputs, with 2
selectable ranges
Digital I/O
16 Digital Inputs (TTL)
8 Solenoid Driver Outputs
10 Digital Reed Relay Contact Outputs
Serial Ports
1 RS-232/485 port
1 RS-485 External Accessory port
Other Ports
3 Full Speed USB ports (one in front, two in rear)
1 Gigabit Ethernet port
Communication Protocols
MODBUS, Streaming
Approvals and Certifications
CE, TUV-SUD Safety
1
Gas Composition of span gas (as defined by user) is incorporated by the MFCs to generate precise
Calibration gases for CO, CO
2
, NO, NO
2
, SO
2
, and others.
Summary of Contents for 146iQ
Page 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Page 2: ......
Page 8: ......
Page 22: ......
Page 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Page 160: ......
Page 170: ......
Page 176: ......
Page 252: ......
Page 300: ......
Page 301: ......