Servicing
MFC Replacement
Thermo Scientific
146iQ Instruction Manual
7-59
Figure 7–66.
MFC Replacement, Plumbing
6.
Using a 9/64 hex drive, unfasten two 8-32 soc cap head screws.
7.
Pull MFC upwards.
8.
Replace MFC and assemble in reverse order.
Note
Orientation of MFC flow arrow facing towards captive hardware (in
unit flow arrow facing towards touchscreen).
▲
Figure 7–67.
MFC Replacement