![ThermoFisher Scientific 146iQ Instruction Manual Download Page 271](http://html1.mh-extra.com/html/thermofisher-scientific/146iq/146iq_instruction-manual_1105634271.webp)
Optional Equipment
Connecting External Devices
Thermo Scientific
146iQ Instruction Manual
9-11
The digital I/O board consists of:
●
16 Digital Inputs (18 pin connector)
●
10 Digital Relay Switches (20 pin connector)
●
8 Valve Driver Outputs (16 pin connector)
The digital inputs are TTL (3 V or 5 V) compatible and are pulled high
within the instrument. The active state can be user defined in firmware.
●
Logic Low Threshold: 0.8 V
●
Logic High Threshold: 2.0 V
●
Absolute allowable input voltages: -0.5 to 5.5 V
Figure 9–6.
Digital Inputs
Digital I/O Board
Digital Inputs
Summary of Contents for 146iQ
Page 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Page 2: ......
Page 8: ......
Page 22: ......
Page 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Page 160: ......
Page 170: ......
Page 176: ......
Page 252: ......
Page 300: ......
Page 301: ......