![ThermoFisher Scientific 146iQ Instruction Manual Download Page 268](http://html1.mh-extra.com/html/thermofisher-scientific/146iq/146iq_instruction-manual_1105634268.webp)
Optional Equipment
Connecting External Devices
9-8
146iQ Instruction Manual
Thermo Scientific
4.
There are six columns for each of the six available output channels:
●
Output (V):
Displays the actual output level at the terminal of the
analog output board. For analog voltage, this value will default at zero.
For analog current, this value will default at 4 mA.
●
Decrease
,
Decrease
,
and Decrease
:
Decreases the output by
coarse, medium, or fine amounts.
●
Increase
, Increase
, and
Increase
:
Increases the output by
coarse, medium, or fine amounts.
●
Commit:
Accepts the changes to the analog output levels.
5.
For the desired analog output channel, increase or decrease the output
until the reading on the recording device indicates the proper value.
6.
After making changes to the output levels, the commit button will turn
green. To accept the changes, press the Commit button. To revert to
the previous values, press the back button to return to the previous
analog output calibration screen.
Summary of Contents for 146iQ
Page 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Page 2: ......
Page 8: ......
Page 22: ......
Page 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Page 160: ......
Page 170: ......
Page 176: ......
Page 252: ......
Page 300: ......
Page 301: ......