The Compensation screen provides compensation for any changes to the
instrument’s photometer output signal due to internal instrument
temperature and pressure variations.
Home Screen>Settings>Photometer Settings>Advanced Measurement
Settings>Compensation
The Compensation screen contains the following information:
●
Temp Compensation:
Toggles temperature compensation enabled or
disabled and provides compensation for any changes to the
instrument's photometer output signal due to internal instrument
temperature variations. The effects of internal instrument temperature
changes on the analyzer's subsystems and output have been empirically
determined. This empirical data is used to compensate for any changes
in temperature.
●
Pressure Compensation:
Toggles pressure compensation enabled or
disabled and provides compensation for any changes to the
instrument's photometer output signal due to bench pressure
variations. The effects of bench pressure changes on the analyzer's
subsystems and output have been empirically determined. This
empirical data is used to compensate for any change in bench pressure.