Servicing
Optional DMC Ozonator
Thermo Scientific
146iQ Instruction Manual
7-45
Use the following procedure to replace the ozonator DMC board.
Equipment required:
Phillips drive, #2
1.
Remove the ozonator DMC from the instrument. Refer to “
DMC
Ozonator Removal
” on page
7-41
.
2.
Unplug heater and lamp cable.
3.
Using a #2 Phillips drive, unfasten five #6-32 pan head screws.
Figure 7–51.
Replacing the Ozonator DMC Board
4.
Replace ozonator DMC board and reassemble in reverse order.
Ozonator DMC
Board Replacement