Operation
Settings
Thermo Scientific
146iQ Instruction Manual
3-51
The Flow and Pressure screen provides status and alarms related to the flow
and pressure module. If an item being monitored goes outside the lower or
higher alarm limit, an alarm is activated.
Use the
and
buttons to move up and down and the
and
buttons to move left and right.
Home Screen>Settings>Health Check>Status and Alarms>Flow and
Pressure
The Flow and Pressure screen contains the following information:
●
Across:
●
Flow and Pressure:
This column lists items associated with the flow
and pressure module.
●
Value:
Displays the current value for each item.
●
Low Alarm:
Displays low alarm status for each item.
●
High Alarm:
Displays high alarm status for each item.
●
Units:
Displays units for each item.
●
Down:
●
Photometer Flow:
Displays the flow reading through photometer.
User can adjust low and high alarm limits.
●
Ozonator Flow:
Displays the flow reading through ozonator.
●
Ozonator Diff Pressure:
Displays differential pressure across the
capillary restricting flow to the ozonator.
●
Photometer Pressure:
Displays photometer bench pressure. User can
adjust low and high alarm limit.
●
Pump Pressure:
Displays the current pump pressure reading.
Flow and Pressure
Summary of Contents for 146iQ
Page 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Page 2: ......
Page 8: ......
Page 22: ......
Page 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Page 160: ......
Page 170: ......
Page 176: ......
Page 252: ......
Page 300: ......
Page 301: ......