Operation
Settings
Thermo Scientific
146iQ Instruction Manual
3-83
The GPT Setup screen allows the user to set up to 6 NO
2
span
concentrations, the NO span concentration targets for each of the specified
NO
2
span concentrations, and total flow. The minimum and maximum
allowable values are calculated based on setpoints, NO tank concentration,
achievable ozonator outputs, and acceptable MFC operation ranges.
Note
The NO tank to be used for GPT should be connected to Solenoid
6. Additionally, NO has to be selected as the gas name in the Cylinder
Setup screen (do not use user-defined names for this tank).
▲
Home Screen>Settings>Ozonator Setup>GPT Setup
Home Screen>Settings>Ozonator Setup>GPT Setup>Continue
The GPT Setup screen contains the following information:
●
Across:
●
NO2 Output:
This column lists items associated with the NO
2
output.
GPT Setup