Operation
Calibration
3-10
146iQ Instruction Manual
Thermo Scientific
The Ozonator Calibration screen allows the user to run automated
calibrations for the ozonator. The automated calibration can only be
performed if the photometer option is enabled. It is recommended to
calibrate the photometer prior to proceeding to the ozonator calibration.
Use the
and
buttons to move up and down and the
and
buttons to move left and right.
Home Screen>Calibration>Ozonator Calibration
Home Screen>Calibration>MFC Calibration>Continue to Ozonator
Lamp Power to Concentration
The Continue to Ozonator Lamp Power to Concentration screen contains
the following information:
●
Across:
●
Description:
Lists items in table.
●
Lamp Power to Concentration:
User can initiate calibration by
pressing Start, and end calibration prematurely by pressing Stop.
Ozonator Calibration