The Select Events 1
15 and 16
31 screens allow the user to select up to 31
events for each program.
Use the
and
buttons to move up and down and the
and
buttons to move left and right.
Home Screen>Settings>Programmed Events>Select Events 1
15
The Select Events screen contains the following information:
●
Across:
●
Program Events:
This column lists event numbers.
●
Action:
User enables/disables event.
●
Duration (min):
Time duration for the event.
●
Solenoid:
User chooses gas solenoid (includes GPT NO
2
, O
3
, and
permeation “solenoids”) or off.
●
Zero/Span:
User chooses zero or span 1
6.
●
Photometer:
User chooses Internal, External, Control, or Off.
●
User D.O. 01
16:
User chooses None, Off, or On for digital
outputs.
●
Down:
●
1
15 and 16
31:
User can set up to 31 events for each program.