Installation and Setup
Setup Procedure
2-6
146iQ Instruction Manual
Thermo Scientific
Use the following procedure to setup the instrument:
1.
Connect a source of zero-air to the inlet port labeled ZERO AIR IN.
The zero-air source should be capable of supplying the full-scale flows
of the zero-air controller at a pressure between 10 and 40 psi.
2.
Connect the standard gas cylinders to the ports labeled GAS1
GAS6.
3.
Connect the
EXHAUST bulkhead to a suitable vent. The exhaust line
should be 1/4-inch OD with a minimum ID of 1/8-inch. The length of
the exhaust line should be less than 10 feet. Verify that there is no
restriction in this line.
4.
Connect a suitable recording device to the rear panel connector. For
detailed information about connecting to the instrument, refer to:
“
Connecting External Devices
” on page
9-1
Communications > “
Analog I/O
” on page
3-58
, and “
Digital I/O
”
on page
3-60
.
5.
Plug the instrument into an outlet of the appropriate voltage and
frequency.
The 146iQ is supplied with a three-wire grounding cord. Under no
circumstances should this grounding system be defeated.
▲
Setup Procedure
Icon Here
Summary of Contents for 146iQ
Page 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Page 2: ......
Page 8: ......
Page 22: ......
Page 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Page 160: ......
Page 170: ......
Page 176: ......
Page 252: ......
Page 300: ......
Page 301: ......