![ThermoFisher Scientific 146iQ Instruction Manual Download Page 254](http://html1.mh-extra.com/html/thermofisher-scientific/146iq/146iq_instruction-manual_1105634254.webp)
System Description
Permeation Oven Board Replacement
8-2
146iQ Instruction Manual
Thermo Scientific
●
Step POL board
●
Sample pump (with photometer)
●
Solenoid valve assemblies
●
Flow Pressure DMC with flow restricting capillaries
The 146iQ can be configured with flow controllers for providing
calibration gas to other instruments. These flow controllers regulate the
amount of calibration gas by mixing a flow of span gas using one flow
controller with zero air or a dilution gas using a second flow controller. The
flow controllers are digitally controlled via a RS-485 interface.
Provides controlled amounts of zero air or a dilution gas for diluting span
or calibration gases by a specified amount to provide a known calibration
gas output from the 146iQ to the external analyzer.
A flow controller to provide a known amount of calibration gas to be
diluted.
The 146iQ can be configured with a second span gas flow controller to
provide a greater operating range of calibration gas output from the 146iQ
to the external analyzer.
The optical bench has an airtight chambers that alternately contain the
sample and reference gases with a photometer lamp at one end and a
detector at the other end.
The photo-diode in the detector transmits light intensity information to
the DMC board for sample measurement computations.
The photometer lamp power supply generates high voltage AC and
contains heater control circuits for the photometer lamp.
The optional internal ozonator operates on the photolytic principle. The
ozone level produced is a function of light intensity at 185 nm and gas
flow. The light intensity is varied by changing the current into the lamp.
Flow Controllers
Zero Air/Dilution Gas
Flow Controller
Span Gas Flow Controller
Span Gas Flow Controller
Range 2 (optional)
Photometer DMC
with Lamp
Detector System
Photometer Lamp Power
Supply
Ozonator DMC
Summary of Contents for 146iQ
Page 1: ...146iQ Instruction Manual Multi Gas Calibrator 117633 00 1Sep2019...
Page 2: ......
Page 8: ......
Page 22: ......
Page 37: ...Operation Instrument Display Thermo Scientific 146iQ Instruction Manual 3 7...
Page 160: ......
Page 170: ......
Page 176: ......
Page 252: ......
Page 300: ......
Page 301: ......