MAINTENANCE ADJUSTMENTS
Maintenance—492/492P Service Vol. 1 (SN B030000 & up)
The following procedures are not part of the regular calibration. They are applicable when an assembly is replaced or after
major repair.
110 MHz IF Assembly Return Loss Calibration
NOTE
The IF assembly must be removed to gain access to the adjustments.
Table 4-2
EQUIPMENT REQUIRED
Test Equipment
Characteristics
Recommended Type
Spectrum Analyzer
Frequency range >110 MHz
TEKTRONIX 492/492P or 7L18 Spectrum
Analyzer
Signal Generator
Frequency 110 MHz at +10 dBm
TEKTRONIX SG 503 for the TM 500 Series
VSWR Bridge
Wiltron VSWR Bridge, Model 62BF50
10 dB and 1 dB Step
Attenuators
50 Ω, 0 dB to 40 dB
TEKTRONIX 2701 Attenuator
Termination
50 Ω
Tektronix Part
No.
011-0049-01
Adapter
Bnc-to-Sealectro
Tektronix Part No. 175-0419-00
4
4
4
4
4
4
*
1.
Test equipment setup is shown in Fig. 4-10. Apply
110 MHz at 2 V peak to peak (+10 dBm) through 35 dB
attenuation to the RF Input of the VSWR bridge. Connect
the RF Out of the VSWR bridge to the RF Input of the
spectrum analyzer. (Do not connect the 110 MHz IF to the
VSWR bridge.)
2.
Set the spectrum analyzer center frequency to
110 MHz, SPAN/DIV to 5 MHz, RESOLUTION BAND
WIDTH to 3 MHz, VERTICAL DISPLAY to 10 dB/DIV, and
REFERENCE LEVEL to -2 0 dBm.
3.
Adjust the step attenuator for full screen (-2 0 dBm)
display.
4.
Connect the 110 MHz IF input to the VSWR bridge
and connect a 50 Ω termination to the output of the IF ampli
fier. Now plug the power cable P3045 into the + and — 15 V
source and ground the case of the assembly.
5.
Adjust C2047 and C325 (Fig. 4-11) simultaneously for
j
minimum signal amplitude on the spectrum analyzer display.
REV JUN 1983
Minimum amplitude must be at least 35 dB down from the
full screen reference of —20 dBm.
6
.
Disconnect test equipment setup and replace the
110 MHz IF assembly.
Do not open the assembly. Adjust the tuning slug only
after checking the filter characteristics.
2072 MHz 2ND CONVERTER
The 2nd Converter assembly consists of a four cavity
2072 MHz bandpass filter, mixer, and a 110 MHz lowpass
filter. The assembly is calibrated at the factory, prior to in
stallation, and requires no calibration after it is installed. We
recommend replacing the assembly if it should malfunction.
The following procedure describes adjustments that can be
made if the biasing should malfunction or the seal on any of
the filter tuning slugs is broken. The mixer diodes are not to
be replaced in the field. Return the assembly to Tektronix,
Inc., for repair.
4-19