ag.
Peak the signal with the +829 MHz IF, R1045
adjustment.
ah. Change to the 1.7—5.5 GHz band and center a
3.5 GHz marker on screen after degaussing.
ai.
Peak the 3.5 GHz signal with the —829 MHz IF,
R1049, adjustment.
aj. All bands should now track. The signal peak with the
front-panel PEAKING control should occur with the knob
index marker near the front-panel mark.
19. Phaselock Calibration
The phaselock assembly normally requires calibration
only after some part of the assembly has been repaired or
replaced. Phase noise, produced by the phaselock loop, is
specified for -7 0 dBc or better 3 kHz out from the re
sponse. This should be checked before calibrating the
assembly.
a.
Test equipment setup is shown in Fig. 3-53. Remove
the Phaselock module and the two cover plates so all circuit
test points and adjustments are accessible. Plug the assem
bly on extender boards and into the instrument. Use extend
er cables and adapters to reconnect signal cables to their
respective connector (cable with yellow band to J501, cable
with black band to J502, and cable between J500 and J511
on Phaselock Control board).
Calibration—492/492P Service Vol. 1 (SN B030000 & up)
Adjustment Procedure
b.
Switch the 492/492P POWER on, set the TIME/DIV
to MNL, FREQ SPAN/DIV to 50 kHz, and Phaselock on.
c.
Check Offset Mixer—This part of the procedure is
only required after repair or replacement of the Mixer board.
1.
Connect the Direct Input of a frequency counter to
pin N (Fig. 3-54A) and adjust the counter controls for a
count. Note the frequency.
2.
Connect the counter to pin K and note the
frequency.
3.
Connect the counter to the collector of Q1040 and
note the frequency. Frequency should equal the differ
ence between pins N and R (e.g., 25.080 —25.00 =
80 kHz). Disconnect the counter probe from the collector
of Q1040.
Phase Lock assembly on extenders
2727-66
Fig. 3-53. Test equipm ent setup fo r calibrating Phaselock assembly.
REV SEP 1983
3-77